For supplies to the Russian Federation
Certificates:
- Certificate of conformity ОИАЭ.RU.013(OC).00250, valid till 23.12.2018
- Certificate of measuring tool type approval №32178, valid till 19.12.2018
Purpose:
continuous gamma dose rate measurement
Main features and functions:
- measurement of gamma dose rate, and data display (depending on version) or data output via RS485
- automatic self-testing, alarm signal generation, alarm signal display and alarm output via RS485
- two alarm threshold levels programming (warning and emergency levels)
- alarm of measured value "minor change"
- "dry-contact" signal of threshold value is exceeded
- standard (0...20mA / 4...20mA) analog output for measured data transfer to external systems
- standard (0...20mA / 4...20mA) analog input for external device connection and its measured data consideration in algorithms
- detecting unit can be installed at a distance up to 60 m from processing unit
- unified and replaceable components for all "RADSYS" system devices.
Performance specification
Measurement range | 10-7 ÷ 102 Gy/h |
Energy window | 0,05 ÷ 3,00 MeV |
Operating temperature range | 0 to +55 °С |
Operating pressure range | 84,0 to 106,7 kPa |
Operating humidity range | up to 98 % at +35 °С |
Relative error | 20 % |
Seismic stability | up to 8 points |
Protection index | IP65 |
Dimensions and weight | DU: 168×78 mm; 0,2 kg LPDU: 390х200х190 mm; not more than 8 kg LPU: 326х196х106 mm; not more than 4,5 kg |
Corresponds to 3N class requirements according to NP-001-97 (OPB-88/97) |
Power supply | 220 V |
Power output | 35 VA |
Options | - Power supply 24 V
- Display and buzzer
- Remote detection unit
- Built-in detection unit
- Two two-step or one-step relays
- Collimating protection
|