Tel +7 484 394-97-16, Fax: +7 484 394-97-68
Marketing and Sales Department: +7 484 394-97-16
RPE «RADICO» Ltd
Hand Foot Clothing Monitor RZB-205
Certificates:
- Certificate of measuring tool type approval №37932, valid till 01.01.2015
Purpose:
measurement of clothing, shoes and hand surface contamination with beta-active radionuclide, threshold exceeding alarm
Main features and functions:
- personnel hand and foot contamination measurement
- clothing and small items contamination measurement by means of remote detector
- numerical and graphical measurement results representation, detailed data for every unit detector
- alarm of preset thresholds exceeding or non-exceeding in form of graphical representation of all contaminated areas together with audio and voice messages
- human actions during measurement are accompanied with voice commands
- alarm theshold values are set separatel for every channel
- fixed or automatic measurement time setting
- automatic unit detector sensitivity calibration
- detector operation status, measurement parameters, sensitivity values for each measurement channel, and test data of unit operation modes can be displayed.
- measurement results database recording, measurement protocol printing
- data transfer to electronic data carrier
- unauthorized software access protection
- self-testing
Performance specification
Detectors | Plastic scintillation detectors RFD485 Fibre™ |
Measured beta energy window | 0,10 ÷ 2,50 MeV |
Measured surface beta particle flow density | 1 ÷ 1·104 cm-2 · min-1 |
Detector sensitivity to 90Sr+90Y nuclide beta irradiation | not less than 1,5 pulse·s-1/particle·cm-2·min-1 |
Detector sensitivity to 60Co nuclide beta irradiation | not less than 0,3 pulse·s-1/particle·cm-2·min-1 |
Limiting permissible main error, P=0,95 | ±30 % |
Limiting permissible additional relative error under gamma irradiation, P=0,95 | ±20 % |
Operating temperature range | +5 to +45 °С |
Operating pressure range | 84 to 106,7 kPa |
Operating humidity range | 90 % at +35 °C |
Protection index | IP20 |
Corresponds to 4 class requirements according to NP-001-97 (OPB-88/97) | |
Power supply | 187÷242 V, 49÷51 Hz |
Power consumption | not more than 120 VA |
Average MTBF | not less than 20000 hours |
Dimensions, weight | 580х850х1700 mm; not more than 70 kg |