TLD Irradiator IR-2000
Produced by Mirion Technnologies
IR-2000 irradiator is operated by microcomputer and uses step motor for detector movement during irradiation.
Irradiation is executed by movement of detector-keeping slide by source located in irradiator, thus providing equal dose for every detector.
Certificates:
Certificate of measuring tool type approval FI.C.38.541.A№63963 till 27.10.21 (RADOS personal radiation monitoring complex)
Purpose:
RADOS TLD-system sensitivity calibration
Features and functions:
- system calibration
- system linearity check
- individual sensitivity calibration of thermoluminescent material
Performance specification
Capacity/speed | 200 cards(cassettes)/h depending on the program |
Dose measurement range: | 100 microGy – 100 mGy |
External level | <0,1 microGy/h with source of rated activity |
Source | Strontium-90/Ytrium-90, rated activity 1 mCi |
Programming | specified |
Display | 32-character text display with commands/error messages |
Keyboard | membrane, 16 keys |
Mechanism | microcomputer, progressively-controlled direct current motor with visible slide and cassette sensors position |
Operating temperature range | +5 to +40 °С |
Relative humidity | 20 – 75% (non-condensing) |
Storage temperature | -40 to +70 °С |
Overall dimensions | 570х275х360 mm |
Weight | 30 kg |
Power supply | 220 V (+10, - 15 %), 50 Hz (±5 %) |