Spectrometer facility for gamma-emitting radionuclide volumetric activity measurement in liquid SGM-101
Functions and Features:
- medium sampling according to specified algorithm using sampling unit valves by LaBr3 crystal-based scintillation gamma-detectors
- principal identified radionuclide: I-131
- additionally identified radionuclides: I-132, I-133, I-134, I-135, Cs-137, Na-24, Co-60
- medium gamma-spectrum measurement
- measurement unit background radiation measurement
- measuring chamber rinsing with chemically demineralized water (CDW) and purging with compressed air by means of sampling unit valves
- medium consumption control by means of sampling unit flowmeters
- spectrum processing for I-131 specific activity calculation
- calculated values output to the built-in display or to the upper level, if the unit is connected to computer-aided system
measured values archiving - continuous equipment state control, facility state as a whole and state of individual parts output to the built-in display
- operator's interface for facility parts operability control, including operability of spectrometric circuit
- signaling of emergency or warning preset value actuating
Performance specification
Safety class according to NP-001-97 (OPB-88/97): complies with 3N class requirements | |
Volumetric activity measurement range, Bq/m3 | 3,7x105÷3,7x109 |
Registered energy range, keV | 100÷3000 |
Permissible relative measurement error limits of radionuclide volumetric activity (confidence coefficient 0,95), %, not more | 50 |
Facility readings irregularity during 8 hours, %, not more | 10 |
Additional error caused by supply variation (-15 to +10), %, of nominal value, or temperature variation (+5 to +50) °C, not more, % | 5 |
Energy resolution | 11,5 |
Registration efficiency at total absorption peak for Cs-137 nuclide along 662 keV energy line in specific geometry – the source is in geometric center between detectors, relative units, not less | 0,02 |
Maximum input detector circuit load, not less, kHz | 200 |
Operating mode setting time for cold detector, h, not more | 0,5 |
AC power supply | 220 |
Power consumption, V-A, not more | 200 |
Operating controlled medium parameters |
|
Controlled medium: water | input pressure: 70 kg/cm2 |
Chemically demineralized water | pressure: up to 10 kg/cm2 |
Compressed air | pressure: up to 6 kg/cm2 |
Operating environment | temperature: (+20 to +35) °C |
Operating temperature range, C | (+5 to +50) °C (up to +70 °C during 2 h) |
Operating humidity range | up to 80 % at +35 °C and lower temperature, noncondensing |
Operating pressure range | 84,0÷106,7 kPA |
Overall dimensions (LxWxH), weight: | |
Sampling unit: | |
Sampling rack, not more, mm | 1354x704,5x1962; not more than 301,0 kg |
Pressure compensator, not more, mm | 211x231x3200; not more than 3,5 kg |
Lead shielding, not more, mm | 542x300x447; not more than 161,0 kg |
Protection rack, not more, mm | 600x310x660; not more than 53,0 kg |
Supply nozzle, not more, mm | 87x99x257; not more than 11,0 kg |
Control cabinet units: | |
Control cabinet, not more, mm | 660x914x229; not more than 31,0 kg |
Optical alarm sounder SO-13, not more, mm | 70x275; not more than 0,3 kg |