Tel +7 484 394-97-16, Fax: +7 484 394-97-68
Marketing and Sales Department: +7 484 394-97-16
RPE «RADICO» Ltd
XPR-80
Produced by PREMIUM Analyse
Designation
the ß measurement activity inline at Environment Monitoring, Process Control and Radioprotection
The new ionization chamber XPR-80 is notably innovative by its compactness and its equipment: It has a particles interchangeable filterles, as well as a heating system gas which avoids the condensation phenomenon. The XPR 80 could also integrate also an air flow measurement for the activity calculation.
Performance specification
Response factor | |
Tritium | 5 500 (Bq/m3) / fA |
offset | <0,1 fA |
noice | <0,fA1 |
responce time | <20 sec, to 90 % |
Measurement range | |
Tritium | 2,5 kBq/m3 |
size | Ø 215 mm х 626 mm (excluding electrical connectors) |
Inline Implementation | by mounting flange Ø 215 mm with 8 tapping M8 out of axe on Ø 194 mm |
material | stainless chamber AISI 304L electro polished |
ionization volume | 8 liters |
nominal air flow | 3 to 5 m3/h |
gas filter | interchangeable particles filter |
gas heating system | heating coaxial resistance |
air flow measurement | omnidirectional sensor TSI 75 mm |
Preamplifier ref. DT P-LN 1A7 | |
ionization voltage | 150 VDC |
decades | 4 |
measurement range | from 0.2 fA to 400nA |
resolution | 30 aA |
accuracy | 0,5 % от измеренной величины |
stability | 0,3 fA / year |
temperature | 0,5 fA or 0,1% measurement between -5 & +40°C |
sizes | Ø 60 mm х 160 mm |
weight | 800 g |
power supply | from 9 to 36 VDC |
humidity | from 5 to 95 % rel. |