RE-2000 Reader

Produced by Mirion Technologies

ertificates:

Approval Certificate for Measuring Instruments FI.C.38.541.A№63963 till 27.10.21 (a part of IDC RADOS)

Designation
determining of radiation dose, cumulated by TLD-dosimeter

RE-2000 reader produces automatically reading of the light sum, cumulated by dosimeters.

Performance specification

Capacity per loading

200 personnel dosimeters
20 personnel dosimeters - manually

Dynamic range

7 decades

Reference light: short term stability

<0.5 %

High voltage

negligible while using photon counting

Linearity

<1 % deviation

Dark current

<1 μSv 137Cs

Dark current equivalent

negligible while using photon counting

System stability: dose:

< 1 μSv

Считывание паспорта

оптический считыватель просечек двоичного кода

Heating time

<15 min

Processing speed

50 four elements cards per hour
100 two elements cards per hour
180 single elements cards per hour

Heating method

Contactless hot nitrogen

Preheat temperature

read out temperature

Preheat time

0 – 70 sec

Acquisition time

0 – 70 sec

Read out temperature

Up to 400 °C

Post heat temperature

read out temperature

Post heat time

0 – 70 sec