Gamma-radiation spectrometer SKG-01 "Measurement WBC"

Certificates:
Certificate of measuring tool type approval
№39353, valid till 01.04.2015
Procedure: Metrological certificate MVI №40090.1Ж582 dated 24.06.2011

Purpose:
Co-60 and other nuclide activity measurement in human lungs and body

Main features and functions:

  • measurement during routine NPP personnel monitoring in case if preliminary measurement using gamma-radiation spectrometer SKG-AT316A "Control BRS" indicated increased control level of total radionuclide content in lungs
  • semiconductor detector is located under comfortable day-bed for positioning and measurement various body parts
  • determines presence of  51Cr, 54Mn, 59Fe, 58Co, 60Co, 65Zn, 95Zr, 95Nb, 103Ru, 110mAg, 124Sb, 141Ce, 144Ce in lungs and measures their content at the moment of measurement
  • determines presence of 134Cs, 137Cs and 22Na in human body on the assumption that their uniform distribution along the body

Performance specification

Registered gamma-ray quantum energy range

50 ÷ 3000 keV   

Operating temperature range

+10 to +35 °С

Operating pressure range

84,0 to 106,7 kPа 

Operating humidity range

up to 75 % at +30 °С

Corresponds to 4 class requirements according to NP-001-97 (OPB-88/97)

Dimensions, weight

2890х1320х720 mm; not more than 220 kg

Protection index

IP40

Power supply

220 V

Power consumption

150 VA

Options

      • power supply 24 V
      • semiconductor gamma-radiation detector with electric cooler CFG-X-COOL-II-230
      • semiconductor gamma-radiation detector with cooler based on Dewar flask