Dose Calibrator Model VDC 505
Produced by Veenstra Instruments
Dose Calibrator Model VDC 505 (.pdf)
This state of the art dose-calibrator combines the best of two worlds. It has the extensive functionality of a software driven device like the Veenstra VDC-405. It has the robustness of firmware like the Veenstra VDC-404. So if you do not want to introduce another PC in your department or Hotlab, the VDC-505 is the bestalternative.
All VDC dose calibrators come as standard with one well liner and dipper. Optional extras available are:
- Extra lead shielding
- Mo-99 breakthrough set
- Copper dipper
- Extra well liner
- Extra dipper
- Label printer
In terms of connectivity the VDC-505 is state of the art. The dose calibrator can be integrated in a network (Ethernet base) and connected to a PC or extra display. Furthermore the system supports a range of label-printers In terms of functionality the VDC-505 is well equipped for both vial and syringe measurements, including future dose calculations. Furthermore this model offers a complete set of calibrator quality control routines and protocols for daily and quarterly checks. These QC reports can then be downloaded to a PC by a standard RS-232 or even the optional Ethernet interface.
Ionisation chamber | Pressurised (12 bar argon) |
Ionisation voltage | 150 V lithium battery |
Well size | 60 mm Ø x 250 mm |
Saturation | > 200 GBq (Tc-99m) |
Resolution | 0.001 MBq |
Energy range | 25 keV - 3 MeV |
Lead shielding | 3 mm Pb |
Linearity | ± 1 % between 1 MBq and 200 GBq (Tc-99m) |
Electrometer accuracy | ± 1 % |
Battery test accuracy | ± 5 % |
Temperature coefficient | 0.1%/°C between 10°C and 40°C at 5 MBq and up |
Reproducibility | ± 1% over 24 hours, stable conditions |
Overall accuracy | ± 3 % dependant of specific calibration source and geometric variations |
Response time | Maximum 2 seconds for 95% of the end value |
Ranges | Three ranges (auto ranging) |
Isotope factor | Digital adjustment (library over 50 isotopes available) |
Bias correction | Digital adjustment |
Calibration | Digital adjustment |
Zero adjustment | Digital control |
Background subtraction | Digital control |
Battery test | Digital control |
Interface | RS-232C |
Power supply | 5 VDC, 250 mA |
Cable | 2.5 meters |
Dimensions | 150 mm Ø x 420 mm |
Weight | 15.5 kg |