Iodine Monitor UDI-201
For supplies to the Russian Federation
Certificates:
- Certificate of conformity РОСС RU.0001.01АЭ00.40.10.2254, valid till 23.11.2015
- Certificate of measuring tool type approval №31715, valid till 22.11.2018
Procedure: Metrological certificate MVI №45540.8U136 dated 03.07.2008
Purpose:
continuous measurement of volumetric activity of I-131 radionuclide in both molecular and organic form
Main features and functions:
- measurement of volumetric activity of I-131 radionuclide in air in both molecular and organic form; measurement data display (depending on version) and data output via RS485
- work in spectrometer mode
- automatic spectrum stabilization against temperature changes
- automatic self-testing, alarm signal generation, alarm signal display and alarm output via RS485
- two alarm threshold levels programming (warning and emergency levels)
- alarm of measured value "minor change"
- "dry-contact" signal of threshold value is exceeded
- standard (0...20mA / 4...20mA) analog output for measured data transfer to external systems
- standard (0...20mA / 4...20mA) analog input for external device connection and its measured data consideration in algorithms
- pumped air monitoring and consumption measurement
- unified and replaceable components for all "RADSYS" system devices
Performance specification
Measurement range | lower limit: 3,7 Bq/m3, for measurements lasting 0÷15 min. |
Energy window | 0,1 ÷ 3,00 MeV |
Operating temperature range | 0 to +55 °С |
Operating pressure range | 84,0 to 106,7 kPa |
Operating humidity range | up to 98 % at +35 °С |
Relative error | not more than 50 % |
Seismic stability | up to 8 points |
Protection index | IP65 |
Dimensions and weight | DU: 380x300x390 mm; not more than 130 kg |
Corresponds to 3N class requirements according to NP-001-97 (OPB-88/97) | |
Power supply | 220 V |
Power output | 50 VA |
Options | Power supply 24 V |