Spectrum analysis monitor for measurement of gamma-emitting radionuclide volumetric activity in gas SGG-01
Certificates:Certificate of measuring tool type approval №37425, valid till 20.02.2020
Purpose:
computer-aided measurement of gamma-emitting radionuclide volumetric activity in NPP's process gaseous medium, in rooms and ventilation system, gaseous environmental discharge
Main features and functions:
- medium sampling by means of sampling system valves in compliance with specified algorithm
- instrument analysis of monitored medium gamma spectrum
- calculation of discovered radionuclide activity and related values (minimum detectable activity, measurement error)
- determination of measuring unit pollution level
- measuring unit blow-down
- measurement quality control by means of monitored medium sampling (consumption) analysis and measurement equipment operability
- calculated activity values comparison with set thresholds (warning and emergency thresholds)
- visual and sound alarm of threshold exceeding
- automatic (cyclic) measurements, and measurements according to specified schedule
- instrument spectrum storage (for the last month) and process results storage for the complete period of operation
- immediate display of last measurement results
- variation diagrams of monitored parameters for any period of operation
- local computer network connection for remote spectrometer analysis monitor operation
Performance specification
Operating temperature range | +5 to +50 °С |
Operating pressure range | 84 to 106,7 kPa |
Operating humidity range | 95 % at +30 °C |
Protection index | IP22 |
Corresponds to 4 class requirements according to NP-001-97 (OPB-88/97) | |
Power supply | ~220 V |
Dimensions, weight | 940х763х680 mm; not more than 1850 kg |
Registered gamma radiation energy range | 50 ÷ 3000 keV |
Volumetric activity measurement range | 3×102 ÷ 5×109 Bq/m |
Limiting permissible main error (confidence probability 0,95) | not more than 50 % |
Integral nonlinearity of spectrometer scale in measured gamma-ray quantum energy range | not more than 0,025% |
Sampling system |
|
Measurement chamber |
|