Tel +7 484 394-97-16, Fax: +7 484 394-97-68
Marketing and Sales Department: +7 484 394-97-16
RPE «RADICO» Ltd
Ultra low lewel monitor, model 593
Produced by Overhoff Technology Corporation
Model 593 is sensitive only to tritium. A semipermeable membrane is used to isolate tritium oxide for passage into the instrument measuring system. All other sample constituents, including pollutants, radioisotopes, aerosols or particulates are removed and eliminated from the measurement.
Specifically designed to measure ultra low levels of tritium for compliance with 10CFR61 regulations, Model 593 measures to levels of 3x10-9Ci/m3, 100 Bq/m3
Technical specifications
(the Model 593 employs two balanced proportional gas flow counter tubes together with a diffusion permeation system for rapid ultra low level tritium specific measurements. The following specifications apply to the standard unit)
RANGE |
0 to 2 MBq/m3 (0 to 50 μCi/m3) |
DISPLAY | 4 1/2 digit panel meter, in steps of 100 Bq/m3 |
RESOLUTION and NOISE LEVEL | 100 Bq/m3, 1 S.D. (1σ) |
TIME BASE, MEASUREMENT |
10 minutes, pulses are accumulated (measurement repetion rate) over a 10 minute period |
TIME BASE, BACKGROUND | 100 minutes (or 10 minutes) pulses from the back ground compensation counter are averaged over a period as long as 100 minutes to enhance signal stability |
MEASUREMENT,INTERFACE OUTPUTS | 0 - 10 V, linear logarithmic, 1 volt/decade
E to I, 4 - 20 mA, logarithmic 2 mA/decade |
OPTIONAL DISPLAY FEATURES | chart recorder, logarithmic scale |
ALARMS, MEASUREMENT | Single alarm point, logarithmic scale |
ALARMS, SUPERVISORY | sample flow failure |
ALARM INTERFACE | fail safe relay closures |
PROPORTIONAL COUNTERS | Dual copper clad acrylic counter tubes, 2 liter active volume, 2.5 liter wetted volume, 0.001 inch tungsten collector anode |
COUNTER GAS | P - 10 or “MAGIC” gas for high performance Flow rate 250 cc/min, typical |
TRITIUM COLLECTION | transfer via a polysulfone semipermeable diffusion membrane |
SAMPLE FLOW SYSTEM | Oscillating piston pump, flow rate 5 lpm typical |
FLOW METERS | front panel rotameters and LED failure displays for counter gas and sampling system |
ENCLOSURE | 19” relay rack, 48” highj. 24” deep approx, indoor use only |
WEIGHT | less than 200 lbs |
POWER | 115/230V 50/60 Hz, 100 W max. |
ENVIRONMENTAL | temperature +10°C to +30°C |