Ultra low lewel monitor, model 93-WR-T-HTO
Produced by Overhoff Technology Corporation
This wide range tritium monitoring station is designed to meet monitoring requirements established by NESHAP and EPA 40CFR or for DOE establishments for the determination of ultra low levels of airborne tritium.
Limits of resolution attained by this instrument is 10-9 Ci/m3 or even lower. Using linear as well as proportional chamber technology, the upper limit of useful measurement can be extended to 10 Ci/m3 or higher.
Technical specifications
Range | 0 - 10 Ci/m3 |
Sensitivity | 1x10-9 Ci/m3 |
Display | 19" color video terminal showing all current and trend data for measurements, and supervisory information |
Accuracy | +10 %, and/or 10-9 Ci/m3 |
Response Rate | Varies with measurement level, up to 30 minutes at very low levels to 1 second at the upper end of the range |
Alarms | Set point alarms for each channel,supervisory alarms for sample flow counter for flow failure |
High/Low Change Over | Automatic changeover to prevent contamination of the low level measurement system |
Volumes of Chambers | Linear: 2 liters nominal |
Counter Gas Flow | Adjustable from 0.1 lpm to 1 lpm |
Enclosure | Standard 19" relay rack NEMA 12 enclosure, 72" high |